Model 1825 RF Power Sensor Calibration |

|
| Key Features |
- Calibrates 10 MHz to 18 GHz RF power sensors from Agilent (HP), Anritsu, Boonton, Gigatronics, Rohde & Schwarz, TEGAM and others
- Ideal for labs with annual calibration workloads of 25 RF power sensors or less
- Less than 1/3 the cost of a comparable System IIA
- Can calibrate many popular sensors in 30 minutes or less
- The accuracy and reliability you have come to expect from TEGAM
|
- No software to learn
- Compact and easy to use
- All connectors are located on the front panel for easy connection
- Rack mount kit available
- Optional A2LA
|
| The Model 1825 RF Power Sensor Calibrator is an accurate, economical, and reliable instrument for calibrating RF Power Sensors in the 10MHz to 18 GHz frequency range. It is a low-cost, non-automated version of the popular TEGAM System II. |
 |
The Model 1825 includes an internal thermistor power standard complete with heater and a Type IV Bridge Circuit for detection of RF power. DC power levels can be measured to within ±0.003%. A Precision Digital Voltmeter, RF Signal Generator, and Power Meter must be supplied by the user in order to complete the system.
No additional software is required. The requirement to calculate calibration factors can be accomplished with a spreadsheet. The 1825 is not a microprocessor-controlled instrument. Since measurements are taken with an external DVM, the 1825 could be integrated into an automated workstation depending on the DVM used and the customer's ability to provide the appropriate software |